1.Accurate temperature uniformity :
±2.0℃(50~100℃) ±3.0℃(101~200℃) ±4.0℃(101~300℃)
2.Test room air circulation system:
Multi air channels circulation system to provide stable and accurate wind flow on all specimen racks.
3.Thick and reliable thermal insulation layers :
Even if temperature interior is very hot - it can be as high as 200℃, the exterior surface of Oven can be touched by your hand.
Applications
heat process and dry heat test of aviation, automotive, electrical and electronic, materials, energy, chemical, medical and pharmaceutical products.
Standard complied
GB/T 2423.2-2008 ( IEC 60068-2-2:2007)test B: dry heat
MODEL |
CK-AB |
VOLUME/CODE |
50 |
290 |
500 |
600 |
720 |
1000 |
1800 |
CONTROL SYSTEM |
Smart temperature balance system |
TEMPERATURE RANGE |
+50℃~+200℃ |
TEMPERATURE UNIFORMITY |
"±2.0℃(50~100℃)
±3.0℃(101~200℃)
Sensor location to wall distance is 1/10 of chamber interior dimensions.(≧50mm)" |
TEMPERATURE CONSTANCY |
±0.5℃ |
"HEATING TIME
(WHEN EMPTY IN WORK SPACE)"
|
+50℃〜+200℃ about 50min |
INTERIOR DIMENSIONS
(WxHxD)CM |
40×35×35 |
60×95×50 |
80×105×60 |
100×100×60 |
120×100×60 |
140×120×60 |
160×140×80 |
EXTERIOR DIMENSIONS
(WxHxD)CM |
71×76.5×77.5 |
87×166.5×73 |
107×178×83 |
127×175×80 |
151×175.5×84.5 |
171×196.5×85 |
221×214.5×104.5 |
UUT SHLEF NUMBER |
2 |
2 |
3 |
3 |
3 |
3 |
3 |
INTERIOR MATERIAL |
stainless steel #430 |
EXTERIOR MATERIAL |
paint coating steel plate |
INSULATION MATERIAL |
glass wool |
SAFETY DEVICES |
overheat protection, motor overload protection, buzzer, trouble indication, NFB |
ACCESSARIES |
one power cable(2.5m long) |
POWER |
AC 1ф2W+G 220V±10% 50HZ |
AC 3ф4W+G 380V±10% 50HZ |
NOTES:
(1)Ambient: setup room should be well-ventilated, temperature 5〜35℃ and relative humidity ≤75%RH.
(2)Sensor measurement method comply with IEC 60068-3-6:2001/ GB/T 2424.6-2006 / JTM K09 :2009. Parameters measured when sensor location to wall distance is 1/10 of chamber interior dimensions.
(3)Pics and spec for reference only. For more details please contact your sales rep.
|
MODEL |
CK-CB |
VOLUME/CODE |
50 |
290 |
500 |
600 |
CONTROL SYSTEM |
Smart temperature balance system |
TEMPERATURE RANGE |
+50℃~+300℃ |
TEMPERATURE UNIFORMITY |
"±2.0℃(50~100℃)
±3.0℃(101~200℃)
±4.0℃(201~300℃)
Sensor location to wall distance is 1/10 of chamber interior dimensions.(≧50mm)" |
TEMPERATURE CONSTANCY |
±0.5℃ |
"HEATING TIME
(WHEN EMPTY IN WORK SPACE)" |
"+50℃~+200℃ about 50min
+50℃~+300℃ about 90min" |
INTERIOR DIMENSIONS
(WxHxD)CM |
40×35×35 |
60×95×50 |
80×105×60 |
100×100×60 |
EXTERIOR DIMENSIONS
(WxHxD)CM |
71×76.5×77.5 |
93×172.5×79.5 |
113×184×89.5 |
133×181×88 |
UUT SHLEF NUMBER |
2 |
2 |
3 |
3 |
INTERIOR MATERIAL |
stainless steel #430 |
EXTERIOR MATERIAL |
paint coating steel plate |
INSULATION MATERIAL |
glass wool |
SAFETY DEVICES |
overheat protection, motor overload protection, buzzer, trouble indication, NFB |
ACCESSARIES |
one power cable(2.5m long) |
POWER |
AC 1ф2W+G 220V±10% 50HZ |
AC 3ф4W+G 380V±10% 50HZ |
NOTES:
(1)Ambient: setup room should be well-ventilated, temperature 5〜35℃ and relative humidity ≤75%RH.
(2)Sensor measurement method comply with IEC 60068-3-6:2001/ GB/T 2424.6-2006 / JTM K09 :2009. Parameters measured when sensor location to wall distance is 1/10 of chamber interior dimensions.